| TCC1600 (Standard) | TCC1600 (Low Temp Option) | |
| Exterior Dimensions (W × D × H) | 2255 × 820 × 2010 mm | |
| Interior Dimensions (W × D × H) | 1631 × 667 × 1603 mm | |
| Volume | 1743L | |
| Default Lighting Layers | 6 | |
| Shelf Height | 247 mm | |
| Cultivation Area | 4.9 m² | |
| Temperature Range | 3 – 45 °C | -10 – 45 °C |
| Temp. Accuracy | 0.1 °C | |
| Temp. Fluctuation | ≤0.5 °C | |
| Temp. Uniformity | ±2 °C | |
| Default Lighting | LED Cool White (380–780 nm); 300 µmol/m²·s @ 10 cm | |
| Upgrade Lighting Options | 450 nm / 660 nm / 735 nm single; 3-colour; 4-colour; 18-colour; UVA/UVB/UVC | |
| Light Intensity Control | 0–100% (infinitely adjustable per layer) | |
| Air Circulation | Bottom air output, top air return (minimises condensation on culture dishes) | |
| Peak Power (220V) | 4.1 kW | |
| Working Voltage | 220 VAC / 50 Hz | |
| Ambient Temp. Req. | 5 – 35 °C | |
| Control Interface | 10" touchscreen + smartphone APP (PID) | |
| Energy Saving | Patented heat recovery; >50% saving vs. comparable units | |
| Optional Extras | Observation window, bottom air outlet | |
The TCC1600 is Jiupo's 1600L extra-large tissue culture chamber, with a 2255 x 820 x 2010 mm cabinet enclosing a 1631 x 667 x 1603 mm interior, six lighting layers, and 4.9 m² of cultivation area across a 247 mm shelf pitch. Its broad interior is intended for laboratories that process very large batches of plantlets or culture vessels across extended multiplication cycles, offering substantially more shelf space than smaller models without requiring a walk-in room installation. Continuous operation, PID temperature control to 0.1 degrees C accuracy, and the patented bottom-air circulation system are all standard