Ivium D-Module, 2 x 1A/20V/250kHz, 2 Channels

Module

The dModule 1A/20V is a dual channel module for the Ivium-n-Stat which allows ±1A / ±20V compliance (±10V applied) and simultaneous impedance measurements on each channel. IviumSoft allows control of separate channels or all channels simultaneously with a synchronized start. The module is encased in a protective housing for easy handling, and users can upgrade the number of channels in the Ivium-n-Stat in a simple plug and play manner.

Safety

The maximum current or potential of the instrument can be defined by the operator. Thus valuable samples may be protected, and unsafe situations prevented.

Expandibility

Each channel can be expanded with various modules, such as a power booster, multiplexer, etc. See our compatibility table for reference.

Hardware

  • Automatic current range selection over all available current ranges
  • Automatic variable noise filter and potentiostat/galvanostat stability settings
  • Real time data display up to 500 pnts/s. Acquisition up to 300,000 pnts/s is stored in instrument memory
  • Applicable scanrates: 1µV/s to 10,000 V/s
  • Minimum interval time: 3.33µs (0.125µs resolution)
  • Result optimisation by automated tuned filters, gain amplifiers and DC-subtraction
  • User selectable floating operation

Channel Configurations

  • 2 x ±1A / ±20V
SKU: D-MODULE(2X1A) Category:
p

Overview

Description & features of the product

h

Specification

Specification of the product

Product Enquiry

Enquire more about our product

Reviews

Reviews about the product

Overview

Module

The dModule 1A/20V is a dual channel module for the Ivium-n-Stat which allows ±1A / ±20V compliance (±10V applied) and simultaneous impedance measurements on each channel. IviumSoft allows control of separate channels or all channels simultaneously with a synchronized start. The module is encased in a protective housing for easy handling, and users can upgrade the number of channels in the Ivium-n-Stat in a simple plug and play manner.

Safety

The maximum current or potential of the instrument can be defined by the operator. Thus valuable samples may be protected, and unsafe situations prevented.

Expandibility

Each channel can be expanded with various modules, such as a power booster, multiplexer, etc. See our compatibility table for reference.

Hardware

  • Automatic current range selection over all available current ranges
  • Automatic variable noise filter and potentiostat/galvanostat stability settings
  • Real time data display up to 500 pnts/s. Acquisition up to 300,000 pnts/s is stored in instrument memory
  • Applicable scanrates: 1µV/s to 10,000 V/s
  • Minimum interval time: 3.33µs (0.125µs resolution)
  • Result optimisation by automated tuned filters, gain amplifiers and DC-subtraction
  • User selectable floating operation

Channel Configurations

  • 2 x ±1A / ±20V

Specifications

Channel Performance

Number of channels in module 2
Current compliance ±1A
Maximum output voltage ±20V
Maximum applied voltage ±10V
4 Electrodes WE, CE, RE and S (and GND) with 4mm banana plugs
Potentiostat Bandwidth >500 kHz
User selectable response filters High Speed, Standard, and High Stability
Programmable response filter 1 MHz , 100 kHz, 10kHz, 1 kHz , 10 Hz
Signal acquisition dual channel 18 bit ADC, 300,000 samples/s

 

Potentiostat

Applied potential range ±10 V, 0.08mV res
Applied potential accuracy 0.2%, or 2 mV
Current ranges  ±100pA to ±1A
Measured current resolution 0.003% of current range, minimum 3fA – Measured current accuracy : 20pA + 0.025% of FSR

 

Galvanostat

Applied current resolution 0.008% of applied current range
Applied current accuracy 0.2%
Galvanostatic current ranges ±10nA to ±1A
Potential ranges ±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V
Measured poential accuracy 0.2%, or 2mV

 

Impedance Analyzer

Frequency range 10JHz to 250 kHz (optional 10JHz to 1MHz)
Amplitude 0.015mV to 2.0 V, or 0.03% to 100% of current range
DC offset 16 bit dc offset subtraction, and 2 dc-decoupling filters

 

Electrometer

Input impedance >1000 Gohm //<10pF
Input bias current <20pA
Bandwidth >5 MHz

 

Special functions / Dimensions

Ohmic drop compensation 2V/current range, 16 bit resolution, bandwidth 40kHz
Size : w x d x h 3 x 35 x 13 cm
Weight 0.7 kg
PC requirements Windows 8/10/11, with free USB port

Enquire More About Our Product

7 + 15 =